ABB FSM4000-SE41F Operating Instruction page 73

Electromagnetic flowmeter
Hide thumbs Also See for FSM4000-SE41F:
Table of Contents

Advertisement

Submenu/parameter
Totalizer → F
0.0 kg/h
Overflow ← R
0
Totalizer ← R
0
Reset totalizer
Overflow Diff.
0
Totalizer Diff.
0.0 kg/h
Submenu
Current output
Current output
0 - 20 mA
Iout for alarm
Low Alarm
Iout low alarm
3.60 mA
Iout high Alarm
21.80 mA
Error 3 mask
Off
Error 4 mask
Off
D184B140U02
Value range
Comment
Numerical entry
Display or input the forward totalizer value.
0 ... 9999999
Display only
Overflow counter max. 65535. 1 overflow = pulse totalizer
of 9,999,999 units. Resets the display status and
increments the overflow counter by one.
Numerical entry
Display or input the reverse totalizer value.
0 ... 9999999
Select ENTER to reset the forward/reverse flow totalizer
and overflow counters.
Display only
Display the overflow for the diff. totalizer.
Numerical entry
Display or input the diff. totalizer value.
0 ... 9999999
This menu does not appear for the PROFIBUS PA or
FOUNDATION fieldbus.
0 - 20 mA, 4 - 20 mA
Select the current output range.
0 - 10 mA, 2 - 10 mA
For HART protocol always 4 - 20 mA.
Low Alarm
Behavior of the current output during an alarm. Low-
Alarm is available only for current output of 2 - 10 mA and
High Alarm
4 - 20 mA. If error 0 (empty pipe) occurs, the current
output is set to the value programmed for the parameter
"lout during empty pipe" in the submenu "Detector empty
pipe". If error 3 (flowrate > 103 %) occurs, "High Alarm" is
always signaled.
Numerical entry
Low-Alarm is available only for current output of 2 -10 mA
or 4 - 20 mA. Set the current output value during Low
Alarm condition. No selection available if High Alarm is
selected for the current output selection for lout during
alarm and if the transmitter version is standard and the
current output range is
0 - 20 mA
0 - 10 mA
.
Numerical entry
Set the current output value during High Alarm condition.
If error 3 (flowrate > 103 %) occurs, "High Alarm" is
always signaled. No selection available for current output
if Low Alarm is selected for lout during Alarm.
Off, On
Deactivates error 3 (flowrate > 103%), if masking = on
Off, On
Deactivates error 4 (ext. zero return), if masking = on
FSM4000
Parameterization
73

Advertisement