Setting And Verifying Resistance Range - Hioki 3154 Instruction Manual

Digital m? hitester
Table of Contents

Advertisement

18
_____________________________________________________________________
3.3 Setting and Verifying Resistance
Range
Resistance range comes with four types of manual range, 2
(1) Press SHIFT key, then LOCK key when READY status.
(2) Each
(3) When desired resistance range is displayed, press STOP
When manual range is selected and measurement value is
(1) Press SHIFT key, then LOCK key when READY status
(2) After confirming, press STOP key or LOCK key to
______________________________________________________________
Setting Basic Testing Conditions
MΩ, 20 MΩ, 200 MΩ, 2000 MΩ(4000 MΩ at 500/1000
V) and auto range.
Auto range may require some time before measurement
value is displayed because range has to shift immediately
after testing starts. Set resistance range to manual range
according to measuring object to eliminate time to
display.
Setting Resistance Range
Display as shown below appears and lower limit value
indicator blinks and displays one of the following
"AUtO", "2", "20", "200", "2000". ("4000" at 500/1000
V)
AUtO
/
key press switches blinking display.
key or LOCK key to confirm. After determining, returns
to READY status.
over 2000 count, "O.F." (Over flow) is displayed.
When measurement value is less than 190 count,
"U.F." (Under flow) is displayed.
Confirming Resistance Range
to display selected resistance range blinking.
return to READY status.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents