Setting And Verifying Timer; Setting And Verifying Test Duration - Hioki 3154 Instruction Manual

Digital m? hitester
Table of Contents

Advertisement

_____________________________________________________________________

3.4 Setting and Verifying Timer

3.4.1 Setting and Verifying Test Duration

Test duration is the time to apply voltage, measure
resistance and perform PASS/FAIL judgment.
Same Test duration setting is applicable at any test voltage.
When setting at auto range, if the test time is set at a short
time period, there is a possibility that test may finish
without displaying a measurement value because of a range
shift. To prevent this from occurring, please set a test time
longer than 2 seconds (for 25 V or 50 V, 1.5 seconds)
when setting at auto range.
(1) During READY status, press TIMER key and timer
(2) Press
(3) Press STOP key or TIMER key to confirm test duration.
______________________________________________________________
Setting Test duration
indicator starts blinking.
/
key to display desired test duration at timer
indicator.
Hold
/
key down to speed up adjustment of timer
control. (units in [seconds], setting range: 0.5 s to 99 s)
"--" is displayed when test duration is not set.
Indicator automatically switches and displays delay time
setting status and timer display and DELAY lamp start
blinking. To set delay time, see 3.4.2.
Setting Basic Testing Conditions
19