3.2.2 Setting and Verifying Test mode

(1) After selecting lower limit value (other than "----") at
(2) Press
(3) Press STOP key or COMP key to confirm selected
When lower limit value (excluding "----") is set, follow the
instructions below to confirm test mode.
(1) Press COMP key twice when READY status. Confirm
(2) After confirming, press COMP key to return to READY
Set test mode adjusted to test voltage.
NOTE
When lower limit value is set at "----"(not using comparator),
test mode setting is not available.
Ready status does not display test mode.
______________________________________________________________
Setting Test mode
lower limit value setting status, test mode setting status
automatically appears by pressing either STOP key or
COMP key.
Which ever display previously set, FAIL StOP or
Continue starts blinking.
/
key to switch FAIL StOP/ContinuE.
FAIL STOP mode: FAIL judgment stops testing.
CONTINUE mode: FAIL judgment does not interfere
mode.
After determining selected mode, automatically returns
to READY status and displays selected lower resistance
value.
Confirming Test mode
that designated test mode display starts blinking.
status.
with testing.
(Testing stops by STOP key press
or by time set by timer.)