Example Test Condition Setting - Hioki 3154 Instruction Manual

Digital m? hitester
Table of Contents

Advertisement

26
_____________________________________________________________________

3.6 Example Test Condition Setting

The example shows READY status setting display after
setting is complete before START key is pressed.
Measurement value display decimal position may differ
case by case.
When returning to READY status without changing
previous setting condition, the last measurement value and
PASS/FAIL judgment result are displayed again.
(1) When testing with comparator and timer and judging
Test voltage
Test duration 5 s
Delay time
Test mode, delay time and beep sound setting are not
displayed during READY status.
Start testing with above settings. While timer is ON, FAIL
judgment stops test followed by an approximately two
second continuous beep sound. When test duration timer
ends, short beep sound repeats three times.
When FAIL STOP mode is set, completing test duration
with timer indicates PASS judgment.
Single continuous beep sound indicates FAIL and repeating
beep indicates PASS judgment.
______________________________________________________________
Setting Basic Testing Conditions
PASS/FAIL with beep sound, set as instructed below.
500 V
ON
Lower limit value
Test mode
Judgment beep sound END(see 5.1.1)
1 MΩ
FAIL STOP