_____________________________________________________________________

3.1 Setting Test voltage

Setting at improper voltage may damage equipment to be
CAUTION
tested.
Test voltage can be set by external I/O. In this case, the
3154 test voltage display lamp stays on displaying previous
voltage until testing starts after setting voltage from external
I/O. Do not start the test simply by confirming the voltage
display lamp, as unexpected voltage may be generated.
Start the test only after you have confirmed the test voltage
setting of the external I/O.
Set and select test voltage from 25 V to 1000 V. To set
comparator at each voltage, selecting another voltage
automatically changes lower limit value.
(1) When READY status (with TEST lamp off or blinking),
(2) Press
(3) Press STOP key or VOLT key to complete setting with
During READY status, TEST status, test voltage can be verified
NOTE
by blinking test voltage indicator lamp.
When selecting test voltage with external I/O, test voltage
cannot be set by key press. Set all external I/O VOLT0 to
VOLT2 to LOW or HIGH (not necessary to connect) use key
press for setting.
______________________________________________________________
Setting Basic Testing Conditions
Setting test voltage
press VOLT key to initiate test voltage display lamp to
blink. (Setting)
/
key to select test voltage to start and test
voltage indicator lamp starts blinking.
test voltage indicator lamp no longer blinking but on
and returns to READY status.
Test voltage indicator lamp