Test Flow - Hioki 3154 Instruction Manual

Digital m? hitester
Table of Contents

Advertisement

30
_____________________________________________________________________

4.1 Test Flow

Simple test flow chart
・During READY status, press START key to start testing.
・Switches to TEST status and TEST lamp is ON.
______________________________________________________________
Testing
Delay time
ON
1. Resistance measurement 2. Delay time countdown
1. Resistance measurement 2. PASS/FAIL judgment
3. Test duration countdown
Lower limit
value
OFF
Test
duration
OFF
Press STOP
key.
・Returns to READY status.
Continues to display measurement value and judgment result
(when lower limit value is set).
・For safety, discharge charge buildup in tested equipment.
During discharge, TEST lamp blinks.
OFF
ON
OFF
CONTINUE mode
ON
Designated Test
duration passes.
Ends testing.
FAIL
judgment
ON
Test mode
FAIL STOP
mode