2.3 Connecting Cord and Probes
(9257, 9289, 9294)
To avoid the risk of electric shock, be sure to press the
3154 STOP key and confirm that the TEST lamp is off
WARNING
before connecting/disconnecting probes. Do not keep
power on equipment to be tested when connecting
probes.
Before using the product, make sure that the insulation
on the test probes is undamaged and that no bare
conductors are improperly exposed. Using the product
under such conditions could result in electrocution.
Replace the test specified by Hioki .
For safety reasons, only use the optional CONNECTION
CAUTION
CORD, 9257 and TEST PROBEs, 9289 or 9294, provided
with the unit for measurement.
To avoid damaging the probes, do not bend or pull the probes.
Although measurement terminals are both located in the front
and rear panels, both cannot be used at the same time. Leave
one side disconnected.
The 9257, 9289 and 9294 come with protective caps. Take the
caps off before connecting to 3154.
Connect black probe to LOW terminal and red probe to HIGH
terminal. Tightly insert probes all the way into the terminals.
______________________________________________________________
Preparation for Testing
9289 TEST PROBE
See 8.4 Options and specifications, about the 9257
CONNECTION CORD.
9294 TEST PROBE