During Testing (Test Status) - Hioki 3154 Instruction Manual

Digital m? hitester
Table of Contents

Advertisement

32
_____________________________________________________________________

4.3 During Testing (TEST status)

When switching to TEST status, the 3154 applies test
voltage to equipment to be tested.
1. When delay time is set, starts resistance measurement
2. After passing delay time (or without setting delay time),
Approximately 0.3 s is required after pressing START key to
NOTE
start resistance measurement and delay time countdown.
See 4.4 to display measurement value.
______________________________________________________________
Testing
and delay time countdown.
For details, see 3.4.3 Determining Delay time.
(1) Starts resistance measurement and displays
measurement value. (unit: MΩ)
(2) During delay time countdown, DELAY lamp blinks.
After completing delay time countdown, DELAY
lamp goes OFF.
starts resistance measurement, PASS/FAIL judgment and
test duration countdown.
(1) Starts resistance measurement and displays
measurement value. (unit: MΩ)
(2) Following resistance measurement, performs
PASS/FAIL judgment by comparator.
When measurement value < lower limit value, FAIL
lamp turns ON.
When measurement value
lamp turns ON.
See 5.1.1 to set beep sound for judgment.
When lower limit value is not set, lower limit value
indicator displays "----" without PASS/FAIL
judgment.
(3) Timer indicator indicates test duration countdown.
When test duration is not set, indicator displays "--".
lower limit value, PASS