Terminal Connector Board/Signal Wires - Hioki 3154 Instruction Manual

Digital m? hitester
Table of Contents

Advertisement

48
_____________________________________________________________________

6.1.1 Terminal Connector Board/Signal Wires

1
START
STOP
VOLT0,VOLT1,
VOLT2
TEST
PASS
FAIL
EXT.DCV
______________________________________________________________
External Interface
Pin number
1, 2
3
4
14
5
6
7
8
9
10
11
12
13, 14
Set signal to LOW to enable same function as
pressing 3154 START key. To perform start up
edge detection, it requires more than 60 ms when
set to LOW. Locking on LOW does not release
continuous START signal input. In this case STOP
signal is accepted for input.
Set signal to LOW to enable same function as
pressing 3154 STOP key. To perform start up edge
detection, it requires more than 60 ms when set to
LOW. Locking on LOW does not accept "all test
start" for input.
Combination of these signals enables selection from
six test voltages (including comparator) options or
from seven saved test condition options. See 6.1.5.
During TEST status, selects LOW. Set with options
either to wait until after discharging charge in tested
equipment after completing test or not to wait and
returns to HIGH. See 6.1.6.
Selects LOW when comparator gives PASS
judgment.
Selects LOW when comparator gives FAIL judgment.
Terminal to supply power from external equipment.
Power supply voltage range is 5 to 30 VDC.
Supplies power also by connecting with INT.DCV
terminal.
IN/OUT
Signal name
OUT
INT.GND
IN
EXT.COM
OUT
FAIL
OUT
PASS
OUT
TEST
IN
VOLT2
IN
VOLT1
IN
VOLT0
IN
STOP
IN
START
IN
EXT.DCV
OUT
INT.DCV

Hide quick links: