Patient Data Management Messages - Welch Allyn Connex VSM 6000 series Directions For Use Manual

Connex series
Hide thumbs Also See for Connex VSM 6000 series:
Table of Contents

Advertisement

Directions for use
Symptom
The lamp brightness does not adjust
The handle becomes very hot to the
touch

Patient data management messages

Message
Maximum number of patient
records saved. Oldest record
overwritten.
No data saved.
Patient ID required to save data.
Clinician ID required to save data. The configuration requires a clinician ID to save
Patient ID required to send data.
Patient list is full. Delete some
patients to add more.
Stop intervals to select new
patient.
No connection for send.
Unable to retrieve list.
Unable to identify clinician.
Unable to identify patient.
Touch Clear to delete all data.
Unable to identify clinician.
Possible cause
The platform handle controller PCBA is
defective
The handle assembly is defective
The lamp has been on for an extended
period of time
Possible cause
The maximum number of patient records in the
monitor's memory has been exceeded
No patient data is available
The configuration requires a patient ID to save
data
data
The configuration requires a patient ID to send
data
The maximum number of patients was exceeded Delete a patient from the list to
The monitor is set to take interval readings
No connectivity is available to support sending
data manually or automatically sending data on
manual save
The monitor is unable to retrieve a patient list
from the network
The clinician ID or password is incorrect
The patient ID does not match an ID in the
patient list or network
The clinician ID does not match an ID in the
network
Troubleshooting 215
Suggested action
Call service.
Call service.
Return the handle to the cradle.
Suggested action
On the Review tab, delete old
records to prevent the alarm from
appearing when new records are
saved.
Take or enter vital signs before
saving.
Call for service to modify
advanced settings.
Call for service to modify
advanced settings.
Add a patient ID.
add a new patient.
Stop intervals before changing
the patient.
Call for service to check the
network connection or wireless
settings.
Call for service to check the
network connection or wireless
settings, or to verify that the
server is available.
Confirm the clinician ID and
password (if applicable), and
retry.
Re-enter the patient ID.
To delete all unsaved data, touch
Clear
.
Re-enter the clinician ID.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Connex iws

Table of Contents