46 Troubleshooting

Patient data management messages

Status message
Alarm
Maximum number of patient
records saved. Oldest record
overwritten.
Unable to access patient
information.
Information
No data to save.
Patient ID required to save
data.
Clinician ID required to save
data.
Patient ID required to send
data.
No matching patient ID found.
Patient list is full. Delete some
patients to add more.
Stop intervals to select new
patient.
Possible cause
The maximum number of patient records has
been exceeded
An error occurred when reading the patient
list or patient record during startup
No patient data is available
The configuration requires a patient ID to
save data
The configuration requires a clinician ID to
save data
The configuration requires a patient ID to
send data
The patient ID does not match
The maximum number of patients was
exceeded
The monitor is set to take interval readings
Welch Allyn Vital Signs Monitor 6000 Series™
Suggested action
Go to the Review tab and
delete old records to prevent
the alarm from appearing
when new records are saved.
Power down and restart the
monitor. If the error persists,
call for service.
Take or enter vital signs
before saving.
Require patient
Disable
ID to save readings
Patient IDs
the
available from the
management tab
Advanced settings.
Disable
Require clinician
ID to save readings
the
Clinician IDs
available from the
management
Advanced settings.
Add a patient ID.
Try again with a valid patient
ID.
Delete a patient from the list
to add a new patient.
Stop intervals before
changing the patient.
on
tab,
Data
in
on
tab,
Data
tab in