198 Advanced settings
d. Touch OK.
The General tab will appear.
2. Touch the Parameters tab.
3. Touch the Temperature tab.
4. Specify settings.
Setting
Unit of measure
Display temperature conversion
Default SureTemp Plus site
Anti-theft return timeout
Mode
Enable pulse timer
Enable Celsius only selection
5. Do one of the following:
Specify the manual parameters
The Manual Parameters frame is in the lower right corner of the Home tab. You can
manually enter values for parameters in the frame. In Advanced settings, you can specify
which parameters appear in the frame and enable manual overrides of other parameter
measurements displayed on the device.
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Parameters tab.
3. Touch the Manual tab.
To continue in the Advanced Settings, touch another tab.
To exit the Advanced Settings and return to the Home tab, touch Exit.
Welch Allyn Connex
Action/Description
Select primary units of measure for the temperature
display on the Home tab
Select to display primary units of measure and secondary
units of measure for the temperature display on the
Home tab.
Select the default site for SureTemp measurements. The
default site applies when clinicians power up the
monitor and each time clinicians remove the temperature
probe from the well.
Last site
Select
to set the default to the site selected
for the last measurement.
Select "Disabled" or a timeout value in hours for the
Braun 6000 to lock after being removed from the dock.
None
Technique Compensation
Select
,
Unadjusted
as the Braun 6000 compensation mode.
Technique Compensation improves reading accuracy by
detecting placement of the probe in the ear canal.
Unadjusted places thermometer in mode to detect raw
ear temperature only.
Select to enable pulse timer on the Braun 6000 handle.
Select to enable Celsius only mode, thus disabling the C/
F button on the Braun 6000 handle as well as the
hardware switch.
®