Reset (Pause Or Turn Off) Audio Alarms - Welch Allyn Connex VSM 6000 series Directions For Use Manual

Connex series
Hide thumbs Also See for Connex VSM 6000 series:
Table of Contents

Advertisement

106 Alarms

Reset (pause or turn off) audio alarms

Audio alarm characteristics
After you reset an audio alarm, some tones do not return, but others return after a
pause interval if the condition that caused the alarm persists.
If a new alarm condition occurs during a pause interval, a new audio tone occurs.
If an audio alarm is not paused or turned off after a period of time, a buzzer
accompanies the tone.
Pause or turn off an audio alarm
1. In the Device Status area, touch
2. If multiple alarms are active, a multiple alarm toggle appears in the Device Status
area. The monitor will either autoscroll through the alarm messages, displaying each
for approximately 4 seconds in order of priority, or you can move through the alarms
manually. Respond to multiple alarms manually as follows:
a. Touch
b. Touch
WARNING Do not pause or turn off an audible alarm if patient safety
might be compromised.
Visual indications remain in the parameter frame until the condition is corrected,
the next measurement is taken, or the alarm is cancelled.
In the Device Status area, if the icon changes to
remains, the timer counts down a fixed pause interval of 60 seconds. The alarm
tone returns after the pause interval.
A longer pause interval can be configured in the Advanced tab. You start the
configured pause interval by touching
If you responded to an NIBP alarm and multiple NIBP limits have been
exceeded, the alarm tone pauses, the first message clears, and the next NIBP
limit message appears with a countdown timer. A new NIBP alarm tone sounds
after the countdown unless you touch
limit message.
to pause all audio alarms.
The alarm icon changes to
interval of 60 seconds. The alarm tone returns after the pause interval.
in the Device Status area to view each alarm in the stack.
Welch Allyn Connex
.
and the message
.
to dismiss each remaining NIBP
and the timer counts down a fixed pause
®