Saving vital sign measurements (Continuous Monitoring
profile)
In the Continuous Monitoring profile, measurements are automatically saved as follows:
Continuous measurements (SpO2, SpHb, pulse rate, etCO2, FiCO2, IPI, respiration
rate, and patient movement) are saved once per minute.
When any measurements fall outside of alarm limits, all continuous measurements
are saved.
NIBP and predictive temperature measurements are saved when acquired.
Manual parameters are saved when you touch the OK button on the Manual tab.
Modifiers are set when you complete parameter setup and return to the Home tab.
These modifiers do not appear on the Review tab.
After 24 hours of data has been saved for the selected patient, any data older than 24
hours is deleted from the monitor.
Note

Intervals Monitoring profile

The Intervals Monitoring profile enables you to use alarms and timed NIBP intervals for
patient monitoring.
Note
The monitor does not retain Direct mode temperatures in memory.
Therefore, it is important to note the temperature before removing the
probe from the measurement site and then manually record it in the patient
record.
Patient vital signs data streams to a central station only in the Continuous
Monitoring profile.
Profiles 83