Troubleshooting; Patient Movement Messages - Welch Allyn Connex VSM 6000 series Directions For Use Manual

Connex series
Hide thumbs Also See for Connex VSM 6000 series:
Table of Contents

Advertisement

Troubleshooting

This section presents tables of technical alarm and information messages, as well as
problem descriptions that do not generate messages, to help you troubleshoot issues on
the monitor.
Note
When the monitor detects certain events, a message appears in the Device Status area
at the top of the screen. Message types include the following:
Information messages, which appear on a blue background.
Very low-priority alarms, which appear on a cyan background.
Low- and medium-priority alarms, which appear on an amber background.
High-priority alarms, which appear on a red background.
Technical alarm messages are low or very low priority unless noted in the Message
column.
You can dismiss a message by touching the message on the screen, or, for some
messages, you can wait for the message to time out.
To use these tables, locate the message that displays on the monitor in the left column
of the table. The remainder of the row explains possible causes and suggests actions
that can resolve the issue.
Note

Patient movement messages

Message
Replace the bed sensor.
The bed sensor is disconnected.
Problem descriptions without messages appear at the end of this section.
Instructions to "Call for service" in the following tables mean that you
should contact qualified service personnel in your facility to investigate the
issue.
Possible cause
The sensor is faulty or expired
The cable is faulty or expired
The bed sensor is disconnected from the
monitor
The bed sensor is disconnected from the
extension cable
207
Suggested action
Replace the bed sensor.
Replace the cable.
Check the bed sensor cable to
ensure that connections to the
monitor and the extension cable
are tight.