Welch Allyn Connex VSM 6000 series Directions For Use Manual page 241

Connex series
Hide thumbs Also See for Connex VSM 6000 series:
Table of Contents

Advertisement

Directions for use
Alarm limits
RRa
Patient movement
Alarm delays
Physiological parameter
Pulse rate
SpO2
Motion (EarlySense)
SpO2
SpHb
etCO2
etCO2 no breath detected
Motion low confidence RR/PR
alarm delay
Respiration rate
RRa (Masimo)
CO2
Motion (EarlySense)
1
These delays are configurable in Advanced settings.
2
When the device determines the alarm condition, it generates alarm signals (visual and auditory) and notifies any
system to which it is connected within one second.
Note
Note
Consider ALL of these alarm delays when setting up your monitor and remote alarm system.
Upper limit range of entry
Adult: 5 to 150 breaths per minute. Factory
default: 50 breaths per minute.
Exit sensitivity: 1 to 6. Factory default: 3
EH > 1 minute
Default delay
(seconds)
3
3
10
10
15
Adult: 30
Pediatric: 20
Neonate: 15
180
10
10
180
The maximum alarm signal generation delay of a technical alarm when connected to a central station
is 4 seconds.
There are no inherent alarm condition delays in the system for temperature or blood pressure alarms.
Lower limit range of entry
Adult: 0 to 145 breaths per minute. Factory
default: 3 breaths per minute.
Off
Configurable
Sum of default
range
delays
1
(seconds)
(seconds)
3–120
4
3–120
4
0–30
11
0–10
11
0–30
16
10–60
31
10–60
21
10–60
16
180–1800
181
0–60
11
0–60
11
10–180
181
Specifications 235
Sum of
configurable
delays
2
(seconds)
4–121
4–121
1–31
1–11
1–31
11–61
11–61
11–61
181–1801
1–61
1–61
11–181