Directions for use
NIBP specifications
Pulse rate accuracy (using blood pressure
determination)
Overpressure cutoff
Note
SureTemp Plus temperature module specifications
Units of measure
Temperature range
Temperature measurement range
Less than 37.0 °C
37.0 °C to 39.0 °C
Greater than 39.0 °C
Less than 96.4 °F
96.4 °F to less than 98.0 °F
98.0 °F to 102.0 °F
102.0 °F to 106.0 °F
Greater than 106.0 °F
Braun ThermoScan PRO thermometer specifications (refer to manufacturer's directions for use
for additional information)
Units of measure
Temperature range
Calibration accuracy
Display resolution
±5.0% (±3 beats per minute)
Adult: 300 mmHg ±15 mmHg (40.0 kPa ±2.0 kPa)
Pediatric: 300 mmHg ±15 mmHg (40.0 kPa ±2.0 kPa)
Neonate: 150 mmHg (20.0 kPa) maximum
NIBP performance can be affected by extremes of humidity, temperature, and altitude.
°F, °C; user-selectable
80° F to 110° F (26.7° C to 43.3° C)
Accuracy
±0.2 °C
±0.1 °C
±0.2 °C
±0.4 °F
±0.3 °F
±0.2 °F
±0.3 °F
±0.4 °F
°F, °C; user-selectable
68° F to 108° F (20° C to 42.2° C)
±0.4° F (±0.2° C) for temperatures ranging from 95.9° F to 107.6° F
(35.5° C to 42° C)
±0.5° F (±0.3° C) for temperatures outside of this range
0.1° F or 0.1° C
Specifications 225