Welch Allyn Connex VSM 6000 series Directions For Use Manual page 203

Connex series
Hide thumbs Also See for Connex VSM 6000 series:
Table of Contents

Advertisement

Directions for use
5. (Optional) Change the name of the program:
a. Touch the keyboard icon.
b. Enter the name of the program and touch OK.
6. Specify the settings for this program.
Setting
Readings to average
Delay to start
Time between
Keep if + or -
The settings you selected appear in the Summary area.
7. Do one of the following:
The program is now available for use.
Specify advanced temperature settings
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
To continue in the Advanced Settings, touch another tab.
To exit the Advanced Settings and return to the Home tab, touch Exit.
Advanced settings 197
Action/Description
Select readings to include in the average. You must
select at least two readings.
The last reading you select is the final reading of the
program. For example, if you select Readings 3, 4, and 5,
the program takes five readings.
Unselected readings—in this example, Readings 1 and 2
—are excluded ("discarded") from the average.
Enter the period between the start of the program (the
moment that the "Start intervals" button is selected) and
the start of the first reading.
Enter the period between the end of one reading and the
start of the next reading.
Enter the range that the program uses to establish the
baseline reading. For more information about how this
setting affects the program, refer to the section
"Excluded readings" below.