Maintenance and service

Perform periodic checks

Welch Allyn recommends that each facility conduct periodic checks of each monitor.
1. Check the following at least daily:
2. Visually inspect the following at least weekly:
3. Do the following at least monthly:
Update settings, replace items, or call for service as necessary based on results of visual
inspection. Do not use the monitor if you see any signs of damage. Qualified service
personnel must check any monitor that is damaged for proper operation before putting
the monitor back into use.
Audio, especially at startup
Fan, especially at startup
Touchscreen alignment
Date
Time
The monitor for any damage or contamination
The AC cord for exposed copper
All cables, cords, and connector ends for damage or contamination
All mechanical parts, including covers, for integrity
All safety-related labeling for legibility and adhesion to the monitor
All accessories (cuffs, tubing, probes, sensors) for wear or damage
Documentation for current revision of the monitor
Test the speaker by selecting each alarm volume level and listening for the
different tones
Visually inspect the mobile stand wheels for wear and faulty operation
Visually inspect the mounting screws on the wall or cart for looseness and wear
WARNING Electric shock hazard. Never perform maintenance or service
tasks while the device is connected to a patient.
CAUTION Wall system components shall be replaced by Welch Allyn
service centers or qualified service personnel.