Welch Allyn Connex VSM 6000 series Directions For Use Manual page 165

Connex series
Hide thumbs Also See for Connex VSM 6000 series:
Table of Contents

Advertisement

Directions for use
Note
Note
6. Remove the probe after the temperature measurement is complete and firmly press
the eject button on the top of the probe to release the probe cover.
7. Return the probe to the probe well.
8. Wash your hands to reduce the risk of cross-contamination.
Braun ThermoScan
The thermometer and accessory dock enable you to transfer an ear temperature
measurement to the monitor. The dock also charges the thermometer battery.
Read the thermometer manufacturer's directions for use before attempting to configure,
use, troubleshoot, or maintain the thermometer.
To switch to the Direct mode, touch
Predictive mode measurement is acquired. The temperature
frame displays text indicating that the Direct mode is active.
The monitor sounds a tone to indicate the start of a Direct
measurement. Once you are in the Direct mode, continue to
separate the buttocks and hold the probe in place throughout
the measurement process.
The monitor does not retain Direct mode temperatures in
memory. Therefore, it is important to note the temperature
before removing the probe from the measurement site and
then manually record it in the patient record.
®
PRO thermometer and accessory dock
WARNING Liquids can damage electronics inside the thermometer.
Prevent liquids from spilling on the thermometer. If liquids are spilled on
the thermometer, dry off the thermometer with a clean cloth. Check for
proper operation and accuracy. If liquids possibly entered the thermometer,
remove the thermometer from use until it has been properly dried,
inspected, and tested by qualified service personnel.
WARNING Many environmental variables, including patient physiology and
clinical application, can affect the accuracy and performance of the monitor.
Therefore, you must verify all vital signs information, especially
Temperature, before treating the patient. If there is any question about the
accuracy of a measurement, verify the measurement using another
clinically accepted method. If the thermometer configured with the device
is unavailable for any reason, use a different thermometer.
CAUTION Probe covers are disposable, nonsterilized, and single-use. The
thermometer is also nonsterilized. Do not autoclave the thermometer and
probe covers. Ensure that probe covers are disposed of according to facility
requirements or local regulations.
CAUTION The thermometer has no user-serviceable parts. If service is
required, call your nearest Welch Allyn Customer Service or Technical
Support facility.
Patient monitoring 159
after the