Directions for use
Note
Place the sensor and the NIBP cuff on different limbs to reduce unnecessary alarms
when you monitor these parameters at the same time.
Note
5. Confirm that the monitor displays SpHb or SpHbv data after connection to the
patient.
While SpHb is being measured, the displayed SpO2 and pulse rate are derived from the
same sensor. If SpO2 is not available, the pulse rate is derived from NIBP.
Detaching the sensor during a measurement triggers an alarm.
If SpHb is being measured continuously on a patient for an extended period, change the
sensor location at least every three hours or as indicated by the sensor manufacturer's
instructions.

Pulse rate frame

The Pulse rate frame displays data, information, and controls used in reading pulse rates.
Pulse rate is derived from one of three sensors based on your configuration: the SpO2
sensor, the NIBP sensor, or the EarlySense sensor. The source of Pulse rate
measurements appears in the lower left corner of the frame.
Note
WARNING Patient injury and inaccurate measurement risk.
Do not use tape to secure the sensor to the patient. This can
restrict blood flow, damage the patient's skin, cause
inaccurate readings, and damage the sensor.
WARNING Patient injury risk. Exercise caution whe applying
a sensor to a patient with compromised skin integrity.
Applying tape or pressure to compromised sites can reduce
circulation and cause additional skin damage.
If a sterile sensor is required, select a sensor that has been
validated for sterilization, and follow the sensor
manufacturer's directions for sterilizing the sensor.
A range of sensors is available for different patient sizes and
measurement sites. Consult the sensor manufacturer's
instructions for selecting the correct sensor.
WARNING Patient injury risk. Incorrect sensor application or
excessive duration of sensor use can cause tissue damage.
Inspect the sensor site periodically as directed in the sensor
manufacturer's instructions.
WARNING Inaccurate measurement risk. Pulse rate measurements
generated through the blood pressure cuff or through SpO2 are subject to
artifact and might not be as accurate as heart rate measurements
generated through ECG or through manual palpation.
You can specify pulse tone volume in the Pulse rate vertical tab (located in
the Settings > Setup tab).
Patient monitoring 173