Taking An Spo Measurement; Other Sensors; Troubleshooting; Error Indications And Interpretation - Welch Allyn Spot Vital Signs 420 Series Operator's Manual

Table of Contents

Advertisement

Other Sensors
A wide variety of reusable and disposable pulse oximetry sensors are available for use with the
Welch Allyn Spot Vital Signs. These sensors expand the utility of the pulse oximetry component
of the device.
Warning:
Use only Nellcor brand SpO» sensors and accessories with Welch Allyn's Spot Vital
Signs configured with the Nellcor pulse oximetry module.
Taking an SpO Measurement
To take an SpO2 measurement:
1.
Properly attach the appropriate sensor to the patient.
2.
The pulse signal bar graph illuminates, indicating the relative strength and quality of the
patient's pulses at the sensor site. The sensor takes approximately 10 seconds to determine
the initial SpO2% value and pulse rate. When the initial values are determined, they are
shown in the SpO> display and the Pulse Rate display, respectively.
3.
The SpO2%
and pulse rate are updated approximately every second. The Spot Vital Signs
monitors a patient's SpO? for up to 10 minutes. After 10 minutes, a C9 error code is displayed.
This error code means that the 10-minute time limit has been exceeded.
4.
Removing the SpO2 sensor from the patient ends the measurement period, and the pulse
signal bar graph blanks.
Troubleshooting
Error Indications and Interpretation
The following table of conditions and error codes provides a quick reference of the descriptions
and probable causes of error codes.
For service-level troubleshooting, refer to the service
manual.
Press the Blood Pressure Start/Stop button to reset flashing patient alarm conditions.
General Error Codes
Code
Description
Corrective Action
E11
Internal safety violation
Check patient, contact Technical Service.
C12
Ambient temperature out of range
Adjust ambient temperature or device
location.
C13
Battery failure
Use wall transformer.
E0.0- E9.9 | Temperature module malfunction
Contact Technical Service.
E38
Date and time not set
Set the Date and Time, see "Setting the
Date and Time" on page 27.
44
Welch Allyn

Advertisement

Table of Contents
loading

Table of Contents