Troubleshooting, Maintenance, and Calibration

Troubleshooting: Error Indications and Interpretation

The following table of conditions and error codes provides a quick reference of the descriptions
and probable causes of error codes. For service-level troubleshooting, refer to the service
manual.
Press the white Blood Pressure Start/Stop button to reset flashing patient alarm conditions.
General Error Codes
Code
E11
Internal safety violation
C12
Ambient temperature out of range
C13
Battery failure
E0.0 - E9.9
Temperature module malfunction
E20 - E50
General internal malfunction
Blood Pressure Error Codes
Code
C02
Auto-zero failure
C03
Inflation too rapid
C04
Excessive inflation time
C05
Excessive noise
C06
Measurement was outside of device's
measurement range
E10
Cuff overpressure condition
Temperature Error Codes
Code
C20
Broken/missing probe
P
Loss of tissue contact
Spot Vital Signs
Description
Description
Description
Corrective Action
Check patient, contact Customer Service.
Adjust temperature or device location.
Use wall transformer.
Contact Customer Service.
Contact Customer Service.
Corrective Action
Check for air obstruction, limit patient
movement.
Check for kinked hose or air obstruction.
Check for air leaks.
Check patient condition, cuff placement,
limit patient movement.
Check patient condition.
Check patient condition.
Corrective Action
Replace probe.
Insure proper probe positioning.
39