9. Take sensor and apply a continuous lengthwise 3 mm (1/8-inch) bead of coupling
10. Place sensor in the middle of one of the areas you have cleaned. Ensure that the cable
11. While holding this sensor in-place, bring the chain around the pipe and attach the closest
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
①
Adjusting Nut
②
Bit Snap
③
Spacer Bar
Figure 5-11
Sensor Mounting with E-Z Clamp and Spacer Bar
Note
To surround larger pipes, link multiple lengths of chain together by mating the Bit Snap
and "S" hook.
①
Mate Bit Snap and "S" Hook to extend chain
Figure 5-12
E-Z Clamp S-Hook
compound across the center of the sensor emitting surface.
①
Cable Connector
②
Angled Edge
③
Sensor
Figure 5-13
Sensor
connector is facing away from where the other sensor will be placed.
link onto the hook of the EZ clamp.
④
Sensor
⑤
E-Z Clamp Chain
⑥
Pipe
④
Front Edge
⑤
Emitting Surface
⑥
Coupling Compound
Connecting
5.2 Sensor Wiring
43