7.8 Analog Input Setup

Vo Output Functions
The Vo analog output is a 0-10 VDC signal that varies linearly in relation to a selected
function.
Assigning a function to the voltage output:
1. From the [Analog Out Setup] menu, press <Right Arrow] to access the [Vo1] option list.
2. Move the cursor to the desired data function by pressing <Up/Down Arrow>.
3. To store selection press <ENTER>.
Note
Refer to drawing 1010N-7-7 for Analog output connections.
Pgen Output Functions
The Pgen analog output is a buffered TTL-compliant pulse rate signal that varies linearly
from 0-5000 Hz in relation to a selected data function.
Assigning a function to the Pgen output:
1. From the [Analog Out Setup] menu, press <Right Arrow> to access the [Pgen] option list.
2. Move the cursor to the desired data function by pressing <Up/Down Arrow>.
3. To store selection press <ENTER>.
7.8
Analog Input Setup
Analog Input Setup
Note
Some versions of the FUP1010 portable flow meter have the ability to read analog input
signals. The [Analog Inp Setup] menu cell is not displayed if the operating system of your
model does not include this facility.
The optional Analog Input Setup function assigns an active analog input to a measurement
channel/path. The flow meter provides two DC current (4-20mA) and two voltage (0-10V)
input ports for single channel, Dual Channel and Dual Path units. The DC current input
ranges from a zero level of 4mA to a full scale of 20mA and allows you to enable this port
and then span it to any desired scaling.