Table 10- 3 Hot Key Summary - Siemens SITRANS FUP1010 Operating Instructions Manual

Ultrasonic flowmeters ip67
Hide thumbs Also See for SITRANS FUP1010:
Table of Contents

Advertisement

Table 10- 3
Hot Key Summary
Key
<+>
<->
<Left Arrow>
<Right Arrow>
<CLR>
<ENTER>
<MENU>
<1>
<2>
<3>
<4>
<5>
<6>
<7>
<8>
<0>
<=>
<F1> and <.>
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
Command Line
MinDamp
MaxDamp
TnSet
ZCOSet
HiSet
LoSet
MaskSet
Hold Set
Troubleshooting
10.3 Test Facilities Graph Screen
Description
Expands (magnifies) waveform to view
more detail.
Contracts waveform to view more of
the waveform.
Shifts receive window to the left
(waveform to the right).
Shifts receive window to the right
(waveform to the left).
Brings waveform to the center of the
screen.
Calls up Text Display menu items.
<Left Arrow> to return to graph.
Exits the Graph Screen and returns to
the main menu.
Minimum damping exponent control (+
or - to increase or decrease).
Maximum damping exponent control (+
or - to increase or decrease).
Transit time adjustment (use + or - to
move Tn marker).
Zero Crossover adjustment (use + or -
to move crossover marker).
Signal envelope threshold level (use +
or - to move threshold).
Signal envelope threshold level (use +
or - to move threshold).
Leading edge masking functions (use +
or - to alter number of samples
masked).
Set this number higher if intermittent
mis-registration occurs.
Exits the command line.
Toggle graph between receive
waveform and envelope waveform.
Dumps the digitized waveform data
over the RS-232 port. You must first
leave the Graph Screen mode before
invoking this command.
135