Siemens SITRANS FUP1010 Operating Instructions Manual page 108

Ultrasonic flowmeters ip67
Hide thumbs Also See for SITRANS FUP1010:
Table of Contents

Advertisement

Functions
7.11 Reflexor
Figure 7-3
Connecting Sensor Cables
Connect sensor cables between the sensors and the flow meter. The XMIT (transmit) sensor
is connected to the Up cable connection and RCV (receive) sensor is connected to the Down
cable connection. The UP/DN orientation is important for the in-line sensor configuration. It is
not a factor for the adjacent configuration.
Select Reflexor Operating Mode
Select [Meter Type] as [Reflexor] for the channel that is to be used as Reflexor. Dual channel
units may have either one or both channels used in the Reflexor mode.
106
A
Connect to DN on flow meter for metal
pipe (Receive) / Connect to UP on flow
meter for plastic pipe (Transmit)
Flow Direction
In-Line Sensor Mounting
B
Connect to UP on flow meter for metal
pipe (Receive) / Connect to DN on flow
meter for plastic pipe (Transmit)
Operating Instructions, 03/2012, A5E02951522-02
FUP1010 IP67 Portable

Advertisement

Table of Contents
loading

This manual is also suitable for:

7me3510

Table of Contents