Connecting
5.2 Sensor Wiring
12. Tighten the adjusting nut until the chain is just snug around the pipe. Ensure that the
13. Repeat Steps 7, 8 and 9 for the second sensor leaving it attached to the spacer bar. At
14. Sight along spacer bar to ensure axial alignment to the pipe. Adjust if necessary. Tighten
15. Observing the upstream and downstream orientation, attach the UP (upstream) and DN
16. Proceed to Commissioning (Page 67).
44
chain is straight around the pipe and the sensor contacts the pipe at the white dot just
under the front label. Ensure that there is equal space on either side of the dot between
the edges of the sensor and pipe.
the same time that you place it in the middle of the second prepared pipe area, secure
the spacing bar to the previously mounted sensor by inserting the sensor index screw
through the REF hole on the bar and into lowest hole on sensor (i.e. for those sensors
that have two holes).
both chains and ensure that sensors did not move while tightening. Do not over tighten.
(downstream) cables to the sensors and make snug. Attach the other ends to the UP and
DN terminals of the flow meter (see figure below).
CH 1 Flow Sensor Cable Connectors
To Channel 2 Flow Sensor Set
Figure 5-14
Connecting Sensors to Flow Meter
Channel 1 - Downstream Sensor
Channel 1 - Upstream Sensor
Operating Instructions, 03/2012, A5E02951522-02