5.2.5
Reflect Mount
Reflect Mount using E-Z Clamp and Spacer Bar
The EZ Clamp is a quick and easy way to securely mount sensors on any pipe. The spacer
bar eliminates manual spacing measurements and provides rigidity for mounting the sensors
while maintaining axial alignment.
Before beginning refer to the E-Z Clamp sensor installation diagram example below.
①
②
③
④
Figure 5-10
Ltn Menu Cell
This view only menu cell shows the distance in inches or millimeters between the front faces
of the sensors along the axis of the pipe. If you are mounting the sensors without a track or
spacer bar, you have to space them according to this value. Note that Ltn may be a negative
number for direct mount on very small pipes where the sensor spacing overlaps.
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
Sensor cable connection
1011HPS Series Sensor
Spring Clip
Spacer Bar
Reflect Mount with EZ-Clamp and Spacer Bar
⑤
Customer's Pipe
⑥
Sensor Index Screw
⑦
Sensor Index Screws-Storage Location
⑧
1011PPS Series Sensor
Connecting
5.2 Sensor Wiring
41