Siemens SITRANS FUP1010 Operating Instructions Manual page 21

Ultrasonic flowmeters ip67
Hide thumbs Also See for SITRANS FUP1010:
Table of Contents

Advertisement

2-Path
2-Path flow meters use two measurement channels to achieve a single output via a "virtual"
third channel. The resultant data is the average of the two channels. Only clamp-on or in-line
transit-time operation is allowed. Benefits include highest available precision and enhanced
immunity to distorted flow profile conditions.
WideBeam Transmission
As shown in the figure below, an ultrasonic sensor induces an axial sonic beam within the
wall of the pipe. These vibrations spread along the pipe wall and then enter the liquid in the
form of a WideBeam wave front traveling at an angle to the main pipe axis. The wide beam
"rains" over the receiving sensor. The wide coverage of the receiver is necessary because
the angle of the sonic beam is related to the liquid's sonic propagation velocity by Snell's
Law.
According to this formula, it can be stated that as the liquid sonic propagation velocity
changes so will the angle between the sonic beam and the flow stream.
Therefore, a significant liquid sonic velocity shift could deflect a "narrow" beam transmission
away from the receiving sensor entirely. The upstream vs. downstream transit-time
difference will also be affected by the changing (or refracting) beam angle. This makes it
necessary for systems to continuously compute this angle, since it is subject to varying
degrees of refraction. The flow meter derives the angle by knowing the fixed position of the
sensors, the dimensions of the pipe and the measured transit-time. Therefore, the flow meter
computes the beam angle relative to the axis of the pipe.
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
Sensor Path 1
Sensor Path 2
3.3 Theory of Operation
Average= (Path 1 + Path 2) / 2
Pipe (front view)
Description
19