F4 Reset Procedure - Siemens SITRANS FUP1010 Operating Instructions Manual

Ultrasonic flowmeters ip67
Hide thumbs Also See for SITRANS FUP1010:
Table of Contents

Advertisement

Note
If you receive a Detection Fault message, it is strongly recommended that the Technical
Service Department (http://www.automation.siemens.com/partner) be contacted.
10.2

F4 Reset Procedure

You may encounter an operating problem that blocks access to the Diagnostics Menu, or the
flow meter may operate erratically after exposure to a power transient or some other
traumatic event. These cases may require use of the F4-reset sequence to restore
operation.
The F4-Reset sequence operates on two levels:
● Clear Active Memory
The first F4-Reset deletes all the data currently in Active Memory, but leaves Datalogger
data and all stored Site Setups intact. This is the most desirable method since all you have
to do to restore operation is reload a saved Site Setup.
● Clear All Saved Memory
If the first sequence fails then you have to resort to the second level of the F4 sequence,
which allows you to clear ALL Saved Memory. Be aware that this erases all saved Site
Setups (including flow calibrated sites), Datalogger Data and user-defined pipe and sensor
tables. This will require you to completely re-install the system and repeat all desired default
settings, custom pipe tables, etc. The table below shows the sequence of the [F4] routine:
[Power On/Off + F4]⇒
Clearing only Active Memory
1. Turn off power (if it is currently on). Press <F4> and keep it pressed while you turn on
2. Press <Right Arrow> to access F4 Reset option list. Press <Down Arrow> to switch the
3. To restore operation, press <MENU> to access the installation menu. Create a new site
4. Re-select any Meter Facilities menu items (e.g. RS-232 setup parameters).
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
power. The prompt: [Clr Active Memory? No] appears at the top of the screen.
option list to [Clr Active Memory? Yes]. Press <ENTER> to clear all Active Site Data (but
not saved Site Setups).
setup or recall a stored site setup.
[Clr Active Memory?]⇒
⇑ ⇓
[Clr Saved Data?]⇒
Troubleshooting
10.2 F4 Reset Procedure
⇒No
⇒Yes
⇒No
⇒Yes
125

Advertisement

Table of Contents
loading

This manual is also suitable for:

7me3510

Table of Contents