Connecting
5.2 Sensor Wiring
Installing a 1012T Mounting Track in Direct Mode
The Sensor Installation procedures show how the automatic selection of sensors, mounting
mode and spacing method are established. Examine the figure below, which illustrates a
typical [Install Sensor] menu screen. Note the automatic assignment of model numbers for
the sensor and mounting track, plus the designation of the Number Index.
The combination of two Model 1012TP Mounting Tracks and a spacer guide is the
recommended way to mount sensors in the Direct Mode. This method ensures that sensors
will align exactly 180° from each other and remain spaced the proper distance apart.
The Direct Mount configuration uses a set of two track rail assemblies; one for each sensor,
installed 180° apart on the pipe. The set includes:
● Direct Mode Track Assembly - This track rail has number index holes for inserting an
54
Automatic selection of mounting track part number , mount mode and Number Index
index pin to position the other sensor.