9. Remove battery pack from battery cover.
10. Insert new battery pack into cover making sure the power connector is placed properly in
11. Reconnect battery power connector to P2.
12. Reverse the installation procedure to reassemble.
9.4
Battery disposal
In accordance with EU directive 2006/66/EC, batteries are not to be disposed of using
municipal waste disposal services.
Waste industrial batteries are accepted back by Siemens or by the local Siemens
representative. Please talk to your local Siemens contact
(http://www.siemens.com/processinstrumentation/contacts) or follow the return procedures
(Page 122) of Siemens Flow Instruments.
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
①
Battery P/N 1010-107
②
Underside of battery cover
③
Power Connector J2
Figure 9-3
Battery Removal/Replacement
the cover.
Maintenance and service
9.4 Battery disposal
121