8. Prepare the areas you marked by degreasing the surface, if needed, and removing any
9. Rotate the track assemblies into their original position on the pipe. Use the edge of the
10. This time, before installing each sensor, apply a 3mm (1/8-inch) continuous bead of
11. Once the sensors are in place, secure with its clamping screws. Do not over tighten.
12. Observing the upstream and downstream orientation, attach the UP (upstream) and DN
13. Proceed to Commissioning (Page 67).
5.3

Zero Flow Adjust Menu

Zero Flow Compensation Methods
Unlike turbine flow meters ultrasonic transit-time flow meters provide active flow
measurement right down to zero flow, however, the measurement of the transit-time delta is
dependent on the similarity or "match" of the electronics, cables and ultrasonic sensors.
Consequently some flow offset (or zero offset) may be present in any installation. To
eliminate this residual zero offset Siemens has developed several different methods to
insure proper zero flow compensation. The following paragraphs describe each method and
when they should be used.
AutoZero
When the 1011HP sensors are mounted in the Reflect Mode configuration the AutoZero
routine is automatically invoked at the end of the Initial Makeup. Flow does not have to be
stopped to perform AutoZero since only the pipe wall signal is used in determining the zero
offset and not the liquid component. The AutoZero routine performs a one-time analysis of
the pipe wall component of the ultrasound signal to quantify any residual mismatch in the
hardware. Once the AutoZero routine is complete, the system memorizes this measured
zero offset and subtracts this value from the flow reading.
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
grit, corrosion, rust, loose paint or surface irregularities with the abrasive pipe
conditioning material provided.
Mylar guide as a stop for both tracks and keep them parallel. Align each track with the
"center line" you previously marked on the Spacing Guide. Tighten tracks securely.
couplant compound along the center (the long way) of the contact surface of the sensor.
Also, keep the sensors lifted slightly from the pipe when installing until the sensor is
against the pin; then push down against the pipe.
(downstream) cables to the sensors and make snug. Attach the other ends to the UP and
DN terminals of the transmitter.
Connecting
5.3 Zero Flow Adjust Menu
61