Installing/mounting

4.3 Application Guidelines

4.3
Application Guidelines
Basic Requirements
● Determine pipe material and dimensions.
● Avoid vertical pipes flowing in a downward direction.
● Avoid installation of sensors on the top and bottom of horizontal pipes, if possible.
● Select a location with the longest straight run of pipe.
● Identify upstream piping configuration (elbow, reducer, etc.).
● Pipe surface should be smooth and, if necessary, free of paint.
● Avoid pressure reduction components upstream.
● Avoid mounting on or near weld seams.
● Pipe must be full to achieve proper operation.
4.4

AC Battery Charger

The flow meter has an optional battery charger that operates from a 100, 110 or 220 VAC
(50 Hz or 60 Hz) power source. It has a universal power input that requires no user
switching.
The 1015BCK-1 Power Adapter/Battery charger is NOT waterproof or water resistant. Do
not allow it to become wet, or attempt to use it when wet. Exposure to a wet or submerged
environment will result in damage to the unit that cannot be repaired and could expose the
user to the risk of electric shock.
22
Optional cords can be provided for connection to AC outlets in most countries.
CAUTION
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02