To calculate a voltage output:
1. Set up the multi-meter to read volts, then connect it to the supply and return terminals of
2. Move the highlight to the port to be tested by pressing <Up/Down Arrow>, press <Right
3. Press <ENTER>. This triggers a 2.00 Volts pop-up window. The multi-meter should now
4. If the multi-meter reading does not match, use the numeric keys to type in the multi-meter
5. Press <ENTER> to register setting. This adjusts the meter's DAC (digital-to-analog
6. Re-check the multi-meter to make sure that it is now reading 2.00 Volts.
Pgen Output Trim
Note
Can be trimmed to within 1.25 Hz of nominal.
To calibrate a pulse rate output
1. Connect a frequency counter to the signal and reference terminals of the pulse rate
2. Move the highlight to the port to be tested by pressing <Up/Down Arrow>, press <Right
3. Press <ENTER>. This triggers a 1 kHz pop-up window. The frequency counter should
4. If the frequency counter reading does not match, use the numeric keys to type in the
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
the voltage output under test.
Arrow>, then press <Down Arrow> to move the cursor to [Trim @ 2V].
be reading 2.00 Volts.
reading.
converter) so that a 2.00 Volts output corresponds with 2.00 Volts on the multi-meter.
output under test.
Arrow> and then press <Down Arrow> to move the cursor to [Trim @ 1kHz].
now be reading 1 kHz.
frequency counter reading.
Functions
7.9 Analog Output Trim
99