iPanel software

Insite with ExC
Test connected devices using Webmin
Complete check form
5-18
Perform the following test for iPanel software.
1.
From the Main Menu area, select Data & Pages > iPanel and verify that the
iPanel home page is launched correctly.
2.
Exit the iPanel software.
a.
Select the About button on the iPanel toolbar.
b.
Select Exit iPanel Software on the About iPanel Software window.
Contact your local online support center to find out if they can view the patient
monitor.
Make sure all parameter modules are securely installed in housings to ensure that the
Tram-Rac housing and E-Module Frame is tested as part of the parameter tests.
Connected devices and network communication can be checked using Webmin as
follows:
1.
Log in to Webmin.
2.
Select Information > Device Information.
3.
Verify that all connected module racks, parameter modules, Unity Network ID
interfaces, displays, writers, USB input devices and network connections are
identified.
If a connected device is not found, it is not communicating.
Complete the
"Check form"
CARESCAPE Monitor B850
on page B-3.
2040384-004D