Error messages and codes

Error message
License expired
Configuration changes. Restart required.
Network down
Identical unit & bed name noticed
Identical IP address noticed
Incompatible gas module
Service Monitor - Error Code 0xHOST1001
Service Monitor - Error Code 0xHOST1002
8-8
The following error messages display at the patient monitor if there is a problem with
the monitor.
Possible cause
Trial license expired
Changes to patient monitor settings
No other network device observed on
the MC Network.
A patient monitor with the identical unit
and bed name is on the network
A patient monitor with the identical IP
address is on the network
The module is not compatible.
The temperature is exceeding limits.
CPU voltage is exceeding limits.
CARESCAPE Monitor B850
Action
1. Remove trial license or enter new
activation code.
1. Power cycle the patient monitor.
1. Verify that the patient monitor is
connected to an active network.
Disconnect the patient monitor that
has the identical unit and bed name.
or
Change the unit and bed name of the
duplicate patient monitor unit and bed
name.
Disconnect the patient monitor that
has the identical IP address.
or
Change the IP address of the patient
monitor that has the duplicate IP
address.
Replace the module. See the
"CARESCAPE Monitor B850
Addendum for Device Compatibility".
1. Turn off the patient monitor.
2. Move the patient monitor to a
location with better air flow.
3. If you need further assistance, call
service.
1. Disconnect the patient monitor from
monitoring patients.
2. Back up the clinical and platform
settings.
3. Print the licensing information page
via Webmin.
4. Call service or your local sales
representative for a replacement
board.
2040384-004D