Overview
Visual inspection
5-2
The purpose of this installation checkout procedure is to ensure that the system is
properly installed and configured for use. Record all checkouts performed using the
"Check form"
on page B-3, or equivalent.
The patient monitor and its components should be carefully inspected every 12
months, and each time the equipment is serviced.
Carefully inspect the device for any damage.
Carefully inspect any devices connected to the following ports: M-ports, RS232,
DVI, USB and e-Port/Tramnet.
Carefully inspect all cables connected to all ports, including the MC Network and
IX Network ports.
Verify that the power cord and USB cables are properly secured with the supplied
retaining clips.
Verify that ferrite blocks are installed on all Tram-Net cables. See page 3-20 for
details.