3-20
To connect the Tram-Rac housing to an ePort/Tramnet
connector
1.
Install the ferrite blocks that came with the patient monitor to both ends of the
Tram-Net cables.
a.
Place one end of cable in the groove of the ferrite block, approximately 26
mm (1 in) from the end of the cable.
b.
Press the sides of the block together until you hear a click. Make sure both
snap fingers are fully engaged, and they are flush against the housing.
c.
Repeat steps a and b for the other end of the cable.
d.
Repeat steps a-c for all Tram-Net cables.
2.
Connect the Tram-Rac housing to an ePort/Tramnet connector with ferrite
blocks on the patient monitor.
NOTE
A TRAM module must always occupy the top position in the Tram-Rac
housing. Other modules are installed below it.
When using more than one Tram-Rac housing, one of the Tram-Racs must
have a power supply.
Make sure the Tram-Net cable has ferrite blocks on both ends.
CARESCAPE Monitor B850
Approximately 26 mm (1 in)
ferrite blocks
ferrite blocks
2040384-004D