Non-medical grade displays

Interface devices
2040384-004D
4.
Connect the Type B plug of the standard USB cable to the upstream port (Type B
USB port) on the bottom of the display.
The patient monitor with a non-medical grade display, which is IEC 950-rated or
equivalent, meets UL and IEC specifications if a medical grade isolation transformer
is used. If a non-medical grade display is to be used, the configuration must meet the
IEC 60601-1-1 standard. Refer to IEC 60601-1-1 for requirements if using non-
medical grade displays in the patient environment.
1.
Connect a USB keyboard, mouse, remote, or barcode scanner to available USB
ports.
2.
Connect the following interface devices to an M-Port on the front panel:
Unity Network ID connectivity device
PRN 50-M
keypad
LED
M-Ports M1 - M4
CARESCAPE Monitor B850
DVI-I connector
Type B USB port
Type A USB ports
Type A USB
3-11