Load settings
Activate settings
Software Management
4-28
NOTE
This option is available only when you log into Webmin remotely. The option is
not available locally at the patient monitor.
1.
If you have not already done so,
Configuration tab.
2.
On the Configuration tab, select Settings.
3.
On the Sub-Modules for Settings menu, select Load.
4.
On the Load Settings window, enter the file name or click on Browse to select a
file from the Choose file dialog box.
5.
Select Upload to load the settings.
When the settings are activated, monitors configured for the U.S. will revert to the
international factory default settings. To reload the U.S. factory default settings, use
the U.S. defaults CD and follow the procedure to
NOTE
The patient monitor needs to be in a discharged state.
1.
Log on to Webmin.
2.
Select the Configuration tab.
3.
On the Configuration tab, select Settings.
4.
On the Sub-Modules for Settings menu, select Activate
5.
On the Activate Settings window, select the settings that you want to activate.
6.
Select Submit.
The changes will take effect after the patient monitor is restarted.
For information, refer to the "CARESCAPE Monitor B850 Software Installation
Instructions".
CARESCAPE Monitor B850
"Log into Webmin"
on page 4-4 and select the
"Load settings"
.
on page 4-28.