Networks

3-12
WARNING
MISSED ALARMS—Do not use with iCentral software with
Versions 5.0.2 and earlier or with Mobile Care Server with Version
5.2 and earlier.
WARNING
INCORRECT CALCULATIONS—Using the CARESCAPE
Monitor B850 with the Aware Gateway software version 1.4 or
earlier could result in incorrect patient height and weight
information. This could lead to incorrect drug dose calculations,
hemodynamic calculations, or oxygenation calculations. Prior to
installing the CARESCAPE Monitor B850, please contact the GE
Healthcare Aware Gateway HL7 Integration Engineering Team or
your GE Healthcare service representative to verify or update your
Aware Gateway configuration.
WARNING
EXCESSIVE LEAKAGE CURRENT—Connect only certified UL
60950/IEC 60950 equipment to the Ethernet MC or IX network RJ-
45 connections. This equipment may need to be used with a
separating device, isolation module or redundant grounding in
accordance with the medical system standard IEC/EN 60601-1-1 in
order to meet system leakage current requirements.
NOTE
Refer to the "CARESCAPE Monitor B850 Addendum for Device Compatibility"
for compatible CARESCAPE Network and S/5 Network devices.
1.
Connect the MC Network or S/5 Network cable to the CARESCAPE Network
MC port on the device.
IX Network port
MC or S/5 Network port
2.
Connect an IX Network cable to the CARESCAPE Network IX port on the
device.