Replace or install the third video card
Replace the enclosure (cover, back panel and front bezel)
2040384-004D
1.
Open the unit as described in steps 1 and 2 on page 9-4.
2.
Remove the third DVI-I board.
a.
Remove 4 screws and 4 standoffs.
b.
Remove 2 jackscrews from the third DVI-I board.
Optional DVI-I
board screws
and standoffs
Power
supply
connector
3.
Replace the DVI-I board.
4.
Reassemble the patient monitor in the reverse order of disassembly.
5.
"Replace the cover on the unit"
6.
Complete the procedures in
1.
Open the unit as described in steps 1 and 2 on page 9-4.
2.
Remove 2 screws on front of bezel.
3.
Disconnect light pipe from receptacle on PCB.
4.
Release 3 snap fingers to free bezel from front of chassis.
5.
Remove the bezel.
6.
Replace the bezel.
7.
"Replace the cover on the unit"
8.
Complete the procedures in
CARESCAPE Monitor B850
on page 9-10.
"Recommended checkout"