Connect system components
3-4
WARNING
INCOMPATIBLE DEVICES—Before connecting an interfacing
module to the device, verify compatibility. Verify the connectivity
of device interfaces before using the equipment. Verify the
compatibility of software versions before using the equipment.
WARNING
EXCESSIVE LEAKAGE CURRENT—When interfacing the
device with other equipment, the devices may only be
interconnected with each other or to parts of the system when it has
been determined by qualified biomedical personnel that there is no
danger to the patient, the operator, or the environment as a result.
WARNING
POWER SUPPLY—The device must be connected to a properly
installed power outlet with protective earth contacts only. If the
installation does not provide for a protective earth conductor,
disconnect the monitor from the power line. All devices of a system
must be connected to the same power supply circuit. Devices which
are not connected to the same circuit must be electrically isolated
when operated (electrically isolated RS232 interface).
WARNING
Do not under any circumstances remove the grounding conductor
from the power plug. Always check that power cord and plug are
intact and undamaged.