Speaker - GE CARESCAPE Monitor B850 Technical Manual

Hide thumbs Also See for CARESCAPE Monitor B850:
Table of Contents

Advertisement

Problem
Patient alarms are not
sounding.
Patient data from the patient
monitor does not display at
the other monitor(s)
2040384-004D
Possible cause
The speakers are disconnected.
The speakers are not functioning.
The alarm limits are not set correctly.
The speaker volume is very low.
The patient is not being monitored.
The patient monitor is off the network.
USB cable from display is not connected to
the patient monitor.
A patch cable is not plugged in all the way.
The MC Network is not configured properly.
There is a general networking issue.
CARESCAPE Monitor B850
Action
Make sure the speakers are connected.
Test if the speaker is functioning, using Monitor Setup
> Sound Volumes.
Make sure the alarms are enabled. Consult the user's
manual for correct alarm settings.
Verify that the alarm volume is not turned off. Look in
Alarms Setup > Audible & Visual.
The alarm function remains disabled until the patient is
correctly admitted into the system.
If a patient's waveform and the DISCHARGED
message both display, consult the user's manual for
correct patient admitting instructions.
Verify network connectivity.
Connect the USB cable to the patient monitor.
Verify that the Ethernet cable is securely connected to
the patient monitor.
Try to ping the device from the patient monitor. Refer to
"Ping a TCP/IP network device"
Refer to
"Ping a TCP/IP network device"
on page 8-5.
on page 8-5.