Host asset settings
Modules
Asset settings
4-22
3.
On the National Requirement window, select the applicable option:
Value
None
France
4.
Select Submit.
The national requirements changes take effect after the next system restart.
NOTE
The Host Serial Number field is view only and cannot be changed.
To set the host asset number:
1.
If you have not already done so,
Configuration tab.
2.
On the Configuration tab, select Host Asset Settings.
3.
On the Host Asset Settings window, enter the user-assigned host asset number in
the Change Value to column.
4.
Select Submit.
The host asset changes take effect immediately.
NOTE
This configuration applies only to the PDM.
The Device Serial Number field is view only and cannot be changed.
To set the device asset number of a PDM:
1.
If you have not already done so,
Configuration tab.
2.
On the Configuration tab, select Modules.
3.
On the Sub-Modules for Modules menu, select Assets Settings.
4.
On the Assets Settings window, enter the user-assigned asset number for the
device in the Change Value to column.
CARESCAPE Monitor B850
Description
Select the normal defaults.
Enable the following country specific monitoring:
At least 1 audible and visual occurrence must remind either
as an un-inhibitable audible and visual alarm or as an
audible and recurring recall repeated at least every 3
minutes and also inhibitable.
Heart Rate high alarm limit maximum 230 instead of 250.
Reminder beep will sound every 2 minutes when alarms
have been silenced permanently.