Frame and E-Modules

2040384-004D
The patient monitor provides support to connect multiple parameter modules at a
time. The ePort and Tramnet ports allow connection of E-Module Frame, Tram-Rac
housing and PDM.
Configuration options
See the following table for possible configuration options. For detailed information
on compatibility, refer to the "CARESCAPE Monitor B850 Addendum for Device
Compatibility."
PDM
Tram-Rac
housing
ePort 2/
Tram-Rac
Tramnet 2
housing
connected
(AC Mains
devices
powered)
F5 Frame
F7 Frame
In the table above:
X = supported
Shaded = not supported
WARNING
ELECTRIC SHOCK - Do not use the F7 Frame for standalone
use. Ventilation holes on the F7 E-module Frame will be
covered only if installed within an Aisys, Avance, or Aespire
anesthesia machine.
Example: If a PDM is connected to ePort 2, you cannot connect a second PDM to
ePort 1. However, you can connect one of the following devices to ePort 1:
Tram-Rac housing
Tram-Rac housing (AC Mains powered)
F5 Frame
F7 Frame
CARESCAPE Monitor B850
ePort 1/Tramnet 1 connected devices
Tram-Rac
Tram-Rac
housing
PDM
housing
(AC Mains
powered)
X
X
X
X
X
X
X
X
X
X
X
F5
F7
Frame
Frame
X
X
X
X
X
3-13