Patient monitor functional tests

Displays
Connected devices and configuration information
IX Network
MC Network
7-4
Perform the applicable tests for the devices and features configured for your system.
1.
Power on the patient monitor.
2.
Check that the patient monitor starts up properly and that the display screen
appears without errors.
3.
Configure the patient monitor for the parameters that are connected and check
that they display correctly.
4.
Leave the patient monitor powered on.
1.
Verify that the picture quality is good for all connected displays. Adjust
brightness and contrast if needed.
2.
Verify that the touchscreen is operating properly. Recalibrate, if needed,
following the procedure on page 3-7.
1.
"Log into Webmin"
on page 4-4.
2.
Select Information > Device Information and verify that all connected devices
are displayed.
3.
Select Information > Configuration Information, and verify that all
configuration information is displayed.
If the patient monitor is connected to the IX Network, follow these steps to verify the
connection.
1.
If a laser printer is configured, generate a test page in Webmin. In Webmin,
select Configuration > Printers > Print Test Page.
2.
If Citrix access is configured, launch iPanel. From the patient monitor's main
menu, select Data & Pages > iPanel.
If the patient monitor is connected to the MC Network, follow these steps to verify
the connection.
1.
Check that a network symbol
of the screen.
2.
Verify that another bed can be viewed. From the patient monitor's main menu,
select Data & Pages > Other Patients.
CARESCAPE Monitor B850
is displayed in the upper right corner
2040384-004D